Human Myoglobin Multiplex Immunoassay Kit (ab219032)

Overview

  • Product name
    Human Myoglobin Multiplex Immunoassay Kit
    See all Myoglobin kits
  • Detection method
    Flow cytometry-fluorescent
  • Precision
    Intra-assay
    Sample n Mean SD CV%
    serum 36 3.29%
    Inter-assay
    Sample n Mean SD CV%
    serum 3 3.16%
  • Sample type
    Cell culture supernatant, Urine, Serum, Heparin Plasma, EDTA Plasma, Citrate Plasma, Cerebral Spinal Fluid
  • Sensitivity
    41.4 pg/ml
  • Range
    123 pg/ml - 270000 pg/ml
  • Recovery

    Sample specific recovery
    Sample type Average % Range
    Serum 103 100% - 108%
    Cell culture media 94 91% - 97%
    Heparin Plasma 95 92% - 97%
    EDTA Plasma 83 77% - 91%
    Citrate Plasma 92 82% - 98%

  • Product overview

    Quantify up to 75 proteins and peptides per well, using just 12.5 µL of sample, with FirePlex® immunoassays. The assay takes approximately 3.5 hours, followed by particle analysis using a compatible flow cytometer and data analysis using our free FirePlex Analysis Workbench software.


    Benefits of the assay


    - Quantify both low and high abundance analytes with a typical sensitivity of 0.5 pg /mL and a fully extended dynamic range spanning 4.5 logs


    - Choose from hundreds of analytes to easily build your own custom panel, use a convenient pre-mixed pre-designed panel, or add new analytes to a pre-designed panel


    - Consistent batch-to-batch performance with assays built using the same recombinant monoclonal antibody pairs as our SimpleStep ELISA® and Matched Antibody Pair Kits


    - One FirePlex particle platform for both multiplex miRNA assays and multiplex immunoassays


    Find out more about multiplex immunoassays using FirePlex particles here


    Important - Use of the FirePlex Multiplex Immunoassay Kit also requires the purchase of:


    - a) Human Multiplex Immunoassay Core Kit, ab208203; and


    - b) Multiplex Immunoassay Protein Standard Mix, such as ab219000.


    FirePlex® is a registered trademark in the United States and is an unregistered trademark elsewhere.

  • Notes

    The following information provides the sensitivity range and recommended starting dilutions for validated biological samples.

    Sensitivity range - 28.9 - 55.8 pg/mL

    Recommended Dilution Range
    Serum Plasma-
    Citrate
    Plasma-
    EDTA
    Plasma-
    Heparin

    Cell
    Culture
    Sup

    Urine Saliva CSF SyF Milk BAL
    1:4 1:4 1:4 1:4 1:4 1:4 1:4 1:4 1:4 1:4 1:4
  • Platform
    Flow

Properties

Images

  • Interpolated dilution factor corrected values are plotted (mean +/- SD, n=2).  The mean Myoglobin concentration was determined to be 11.9 ng/mL with a range of 2.9 – 26.3 ng/mL.

  • The concentrations of Myoglobin were interpolated from the Myoglobin standard curves and corrected for sample dilution. 1X diluted samples are as follows:  serum 25%, plasma (EDTA) 25%, plasma (heparin) 25% and plasma (citrate) 25%.  The interpolated dilution factor corrected mean values are plotted.  The mean Myoglobin concentration was determined to be 18.1 ng/mL in serum, 73.0 ng/mL in plasma (EDTA), 17.1 ng/mL in plasma (heparin) and 21.7 ng/mL in plasma (citrate).

  • The concentrations of Myoglobin were interpolated from the Myoglobin standard curve and corrected for sample dilution.  The interpolated dilution factor corrected mean values are plotted.  The mean Myoglobin concentration was determined to be 66126 pg/mL in cell culture media containing 10% fetal bovine serum.

  • The concentrations of Myoglobin were interpolated from the Myoglobin standard curves and corrected for sample dilution. 1X diluted samples are as follows:  25% urine, cerebrospinal fluid (CSF) and bronchial lavage.  The interpolated dilution factor corrected values are plotted.  The mean Myoglobin concentration was determined to be 797 pg/mL in urine, 638 pg/mL in CSF and 45.4 pg/mL in bronchial lavage.

References

ab219032 has not yet been referenced specifically in any publications.

Customer reviews and Q&As

There are currently no Customer reviews or Questions for ab219032.
Please use the links above to contact us or submit feedback about this product.

Please note: All products are "FOR RESEARCH USE ONLY AND ARE NOT INTENDED FOR DIAGNOSTIC OR THERAPEUTIC USE"