JavaScript is disabled in your browser. Please enable JavaScript to view this website.
AB200016

Human Cardiac Troponin I ELISA Kit

5

(2 Reviews)

|

(14 Publications)

Human Cardiac Troponin I ELISA Kit is a single-wash 90-min Simplestep used to quantify Human Cardiac Troponin I with a sensitivity of 4.4 pg/ml. The assay uses a simple mix-wash-read protocol with just one incubation and one wash step.

- Colorimetric Sandwich ELISA - 450 nm readout : works on any standard plate reader
- Design your own immunoassay: we also offer the conjugation-ready antibody pair
- Cited in over 10 citations

View Alternative Names

TNNC1, TNNI3, Cardiac troponin I

13 Images
ELISA - Human Cardiac Troponin I ELISA Kit (AB200016)
  • ELISA

Supplier Data

ELISA - Human Cardiac Troponin I ELISA Kit (AB200016)
Sandwich ELISA - Human Cardiac Troponin I ELISA Kit (AB200016)
  • sELISA

Supplier Data

Sandwich ELISA - Human Cardiac Troponin I ELISA Kit (AB200016)

Interpolated concentrations of native Cardiac Troponin 1 in human heart tissue extract based on a 10 μg/mL extract load.

The concentrations of Cardiac Troponin 1 were measured in duplicate and interpolated from the Cardiac Troponin 1 standard curve and corrected for sample dilution. The interpolated dilution factor corrected values are plotted (mean +/- SD, n=2). The mean Cardiac Troponin 1 concentration was determined to be 1,793 pg/mL in human heart tissue extract.

Sandwich ELISA - Human Cardiac Troponin I ELISA Kit (AB200016)
  • sELISA

Supplier Data

Sandwich ELISA - Human Cardiac Troponin I ELISA Kit (AB200016)

Example of human Cardiac Troponin 1 standard curve in 1X Cell Extraction Buffer PTR.

The Cardiac Troponin 1 standard curve was prepared as described in Section 10. Raw data values are shown in the table. Background-subtracted data values (mean +/- SD) are graphed.

Sandwich ELISA - Human Cardiac Troponin I ELISA Kit (AB200016)
  • sELISA

Supplier Data

Sandwich ELISA - Human Cardiac Troponin I ELISA Kit (AB200016)

Linearity of dilution is determined based on interpolated values from the standard curve. Linearity of dilution defines a sample concentration interval in which interpolated target concentrations are directly proportional to sample dilution.

Recombinant Cardiac Troponin 1 was spiked into the following biological samples and diluted in a 2-fold dilution series in Sample Diluent 50BP for serum and plasma samples and Sample Diluent NS for cell culture media samples.

Neat pooled serum and plasma (EDTA, Heparin, Citrate) samples from healthy donors was measured in duplicate. All values were below the detectable range of the assay.

Sandwich ELISA - Human Cardiac Troponin I ELISA Kit (AB200016)
  • sELISA

Supplier Data

Sandwich ELISA - Human Cardiac Troponin I ELISA Kit (AB200016)

Example of human Cardiac Troponin 1 standard curve in Sample Diluent 50BP.

The Cardiac Troponin 1 standard curve was prepared as described in Section 10. Raw data values are shown in the table. Background-subtracted data values (mean +/- SD) are graphed.

Sandwich ELISA - Human Cardiac Troponin I ELISA Kit (AB200016)
  • sELISA

Supplier Data

Sandwich ELISA - Human Cardiac Troponin I ELISA Kit (AB200016)

Example of human Cardiac Troponin 1 standard curve in Sample Diluent NS.

The Cardiac Troponin 1 standard curve was prepared as described in Section 10. Raw data values are shown in the table. Background-subtracted data values (mean +/- SD) are graphed.

Sandwich ELISA - Human Cardiac Troponin I ELISA Kit (AB200016)
  • sELISA

Supplier Data

Sandwich ELISA - Human Cardiac Troponin I ELISA Kit (AB200016)

Sensitivity.

The MDD was determined by calculating the mean of zero standard replicates and adding 2 standard deviations then extrapolating the corresponding concentration.

Sandwich ELISA - Human Cardiac Troponin I ELISA Kit (AB200016)
  • sELISA

Supplier Data

Sandwich ELISA - Human Cardiac Troponin I ELISA Kit (AB200016)

Comparison of Cardiac Troponin 1 in various human tissue extracts.

Human heart was analyzed based on a 10 μg/mL extract load while human skeletal muscle and human brain extracts were analyzed based on a 200 μg/mL extract load. The concentrations of Cardiac Troponin 1 were measured in duplicate and interpolated from the Cardiac Troponin 1 standard curve and corrected for sample dilution. The interpolated dilution factor corrected values are plotted (mean +/- SD, n=2). The mean Cardiac Troponin 1 concentration was determined to be 182.4 ng/mg in human heart tissue extract and undetectable in human skeletal muscle and brain tissue extracts.

Sandwich ELISA - Human Cardiac Troponin I ELISA Kit (AB200016)
  • sELISA

Supplier Data

Sandwich ELISA - Human Cardiac Troponin I ELISA Kit (AB200016)

Example of human Cardiac Troponin 1 standard curve in 1X Cell Extraction Buffer PTR. The Cardiac Troponin 1 standard curve was prepared as described. Raw data values are shown in the table. Background-subtracted data values (mean +/- SD) are graphed.

Sandwich ELISA - Human Cardiac Troponin I ELISA Kit (AB200016)
  • sELISA

Supplier Data

Sandwich ELISA - Human Cardiac Troponin I ELISA Kit (AB200016)

Linearity of dilution.

Neat serum from ten individual healthy human female donors was measured in duplicate. All values were below the detectable range of the assay.

Native Cardiac Troponin 1 was measured in the following biological samples in a 2-fold dilution series. Sample dilutions are made in 1X Cell Extraction Buffer PTR.

Sandwich ELISA - Human Cardiac Troponin I ELISA Kit (AB200016)
  • sELISA

Supplier Data

Sandwich ELISA - Human Cardiac Troponin I ELISA Kit (AB200016)

Human Cardiac Troponin 1 Standard Curve Comparison between current and original ELISA kit.

Standard Curve comparison between current human Cardiac Troponin 1 SimpleStep ELISA kit and original ELISA kit. The current SimpleStep ELISA kit shows increased sensitivity.

Sandwich ELISA - Human Cardiac Troponin I ELISA Kit (AB200016)
  • sELISA

Supplier Data

Sandwich ELISA - Human Cardiac Troponin I ELISA Kit (AB200016)

Example of human Cardiac Troponin 1 standard curve in Sample Diluent NS. The Cardiac Troponin 1 standard curve was prepared as described. Raw data values are shown in the table. Background-subtracted data values (mean +/- SD) are graphed.

Sandwich ELISA - Human Cardiac Troponin I ELISA Kit (AB200016)
  • sELISA

Supplier Data

Sandwich ELISA - Human Cardiac Troponin I ELISA Kit (AB200016)

Example of human Cardiac Troponin 1 standard curve in Sample Diluent 50BP. The Cardiac Troponin 1 standard curve was prepared as described. Raw data values are shown in the table. Background-subtracted data values (mean +/- SD) are graphed.

Key facts

Detection method

Colorimetric

Sample types

Cell culture extracts, Tissue Extracts, Heparin Plasma, Citrate plasma, Cell culture supernatant, Serum, EDTA Plasma

Reacts with

Human

Assay type

Sandwich (quantitative)

Sensitivity

= 4.4 pg/mL

Range

31.3 - 4000 pg/mL

Assay time

1h 30m

Assay Platform

Microplate (12 x 8 well strips)

Reactivity data

{ "title": "Reactivity Data", "filters": { "stats": ["", "Reactivity", "Dilution Info", "Notes"] }, "values": { "sELISA": { "reactivity":"TESTED_AND_REACTS", "dilution-info":"", "notes":"<p></p>" } } }

Product details

Human Cardiac Troponin I ELISA Kit ab200016 is a rapid single-wash 90-min Sandwich ELISA to measure Human Cardiac Troponin I in cell culture extracts, cell culture supernatant, citrate plasma, EDTA plasma, heparin plasma, serum, tissue extracts. This SimpleStep sensitivity is 4.4 pg/mL.

How the assay works

Human Cardiac Troponin I SimpleStep ELISA®employs capture antibodies conjugated to an affinity tag that is recognized by the monoclonal antibody used to coat our SimpleStep ELISA® plates. This approach to sandwich ELISA allows the formation of the antibody-analyte sandwich complex in a single step, significantly reducing assay time. See the SimpleStep ELISA® protocol summary in the image section for further details.

Assay Specificity

Our SimpleStep ELISA® kits use recombinant monoclonal antibodies rigorously validated to ensure the highest level of consistency and reproducibility, improved sensitivity and specificity and ease of scalability and security of supply.
Please refer to our protocol booklet for more details.

Human Cardiac Troponin I ELISA Kit ab200016 protocol summary

1. Mix: add samples/standards to the wells together with the capture and detector antibody cocktail. Incubate 1 hr at room temperature
2. Wash
3. Add TMB development solution - incubate for 10 min
4. Add Stop solution
5. Read the results on a plate reader at 450 nm

The regulatory troponin complex regulates skeletal and cardiac muscle contraction. This complex, together with tropomyosin, is located on the actin filament and it is composed of three protein subunits: troponin T (the tropomyosin-binding subunit), troponin I (the inhibitory subunit, which inhibits the ATPase activity of acto-myosin), and troponin C (the Ca2+-binding subunit). Troponins T and I have unique cardiac isoforms, whereas cardiac and skeletal muscle share troponin C. Specifically, three human troponin I isoforms have been described: one is expressed in cardiac muscle (Cardiac Troponin I) and the other two are found in slow-twitch and fast-twitch skeletal muscle fibers (slow sTnI and fast sTnI, respectively). The sequence homology between Cardiac Troponin I and slow sTnI is approximately 40% and somewhat less for fast sTnI. Cardiac Troponin I is 209 amino acid long with a molecular weight of approximately 24 kDa. Mouse and rat Cardiac Troponin I proteins both show 93% amino acid identity to human Cardiac Troponin I. The presence of human Cardiac Troponin I in serum (together with chest pain and electrocardiographic changes) is now considered as one highly specific biochemical marker of myocardial injury, risk stratification of acute coronary syndrome and myocardial infarction. Mutations of Cardiac Troponin I are associated with hereditary cardiomyopathy. Specifically, defects in Cardiac Troponin I are the cause of cardiomyopathy familial hypertrophic type 7 (CMH7). Familial hypertrophic cardiomyopathy is a hereditary heart disorder characterized by ventricular hypertrophy, which is usually asymmetric and often involves the interventricular septum. The symptoms include dyspnea, syncope, collapse, palpitations, and chest pain. They can be readily provoked by exercise. The disorder has inter- and intra-familial variability ranging from benign to malignant forms with high risk of cardiac failure and sudden cardiac death. Defects in Cardiac Troponin I also cause cardiomyopathy familial restrictive type 1 (RCM1). RCM1 is a heart muscle disorder characterized by impaired filling of the ventricles with reduced diastolic volume, in the presence of normal or near normal wall thickness and systolic function. Furthermore, cardiomyopathy dilated type 2A (CMD2A) and cardiomyopathy dilated type 1FF (CMD1FF), disorders characterized by ventricular dilation and impaired systolic function resulting in congestive heart failure and arrhythmia, are caused by defects in Cardiac Troponin I.

REACH authorisation
Abcam has not and does not intend to apply for the REACH Authorisation of customers' uses of products that contain European Authorisation list (Annex XIV) substances.
It is the responsibility of our customers to check the necessity of application of REACH Authorisation, and any other relevant authorisations, for their intended uses.

Precision

[ { "reproducibilityType": "Inter", "sample": "Human Heart Extract", "replicates": 3, "mean": null, "standardDeviation": null, "coefficientOfVariability": "7.1" }, { "reproducibilityType": "Intra", "sample": "Human Heart Extract", "replicates": 8, "mean": null, "standardDeviation": null, "coefficientOfVariability": "3" } ]

Recovery

[ { "sample": "Serum", "range": "93 - 101 %", "average": "= 97" }, { "sample": "EDTA Plasma", "range": "90 - 97 %", "average": "= 93" }, { "sample": "Tissue Extracts", "range": "104 - 118 %", "average": "= 112" }, { "sample": "Heparin Plasma", "range": "99 - 117 %", "average": "= 107" }, { "sample": "Citrate plasma", "range": "87 - 100 %", "average": "= 93" }, { "sample": "Cell culture media", "range": "97 - 108 %", "average": "= 102" } ]

What's included?

{ "values": { "1x96Tests": { "sellingSize": "1 x 96 Tests", "publicAssetCode":"ab200016-1x96Tests", "assetComponentDetails": [ { "size":"1 x 6 mL", "name":"Antibody Diluent CPI2", "number":"AB200016-CMP13", "productcode":"" }, { "size":"1 x 20 mL", "name":"Sample Diluent 50BP", "number":"AB200016-CMP02", "productcode":"" }, { "size":"1 x 20 mL", "name":"10X Wash Buffer PT (ab206977)", "number":"AB200016-CMP12", "productcode":"" }, { "size":"2 x 1 Vial", "name":"Human Cardiac Troponin I Lyophilized Recombinant Protein", "number":"AB200016-CMP11", "productcode":"" }, { "size":"1 x 600 µL", "name":"10X Human Cardiac Troponin I Capture Antibody", "number":"AB200016-CMP09", "productcode":"" }, { "size":"1 x 50 mL", "name":"Sample Diluent NS (ab193972)", "number":"AB200016-CMP03", "productcode":"" }, { "size":"1 x 12 mL", "name":"TMB Development Solution", "number":"AB200016-CMP06", "productcode":"" }, { "size":"1 x 12 mL", "name":"Stop Solution", "number":"AB200016-CMP05", "productcode":"" }, { "size":"1 x 1 Unit", "name":"Plate Seals", "number":"AB200016-CMP08", "productcode":"" }, { "size":"1 x 1 Unit", "name":"SimpleStep Pre-Coated 96-Well Microplate (ab206978)", "number":"AB200016-CMP04", "productcode":"" }, { "size":"1 x 10 mL", "name":"5X Cell Extraction Buffer PTR (ab193970)", "number":"AB200016-CMP07", "productcode":"" }, { "size":"1 x 600 µL", "name":"10X Human Cardiac Troponin I Detector Antibody", "number":"AB200016-CMP10", "productcode":"" } ] } } }

Properties and storage information

Shipped at conditions
Blue Ice
Appropriate short-term storage conditions
+4°C
Appropriate long-term storage conditions
+4°C
Storage information
+4°C

Product protocols

Target data

Troponin I is the inhibitory subunit of troponin, the thin filament regulatory complex which confers calcium-sensitivity to striated muscle actomyosin ATPase activity.
See full target information TNNI3

Publications (14)

Recent publications for all applications. Explore the full list and refine your search

Signal transduction and targeted therapy 10:157 PubMed40360476

2025

Exosomal transfer of pro-pyroptotic miR-216a-5p exacerbates anthracycline cardiotoxicity through breast cancer-heart pathological crosstalk.

Applications

Unspecified application

Species

Unspecified reactive species

Yan Ma,Yongjun Wang,Renzheng Chen,Yabin Wang,Yan Fang,Cheng Qin,Tianhu Wang,Xiaoying Shen,Tingwen Zhou,Lei Tian,Ting Sun,Li Fan,Xiaoning Wang,Dong Han,Feng Cao

iScience 27:109139 PubMed38384853

2024

A safety screening platform for individualized cardiotoxicity assessment.

Applications

Unspecified application

Species

Unspecified reactive species

Verena Schwach,Rolf H Slaats,Carla Cofiño-Fabres,Simone A Ten Den,José M Rivera-Arbeláez,Maureen Dannenberg,Chiara van Boheemen,Marcelo C Ribeiro,Sabina Y van der Zanden,Edgar E Nollet,Jolanda van der Velden,Jacques Neefjes,Lu Cao,Robert Passier

ACS sensors 8:3574-3584 PubMed37705448

2023

Paper-Based All-in-One Origami Nanobiosensor for Point-of-Care Detection of Cardiac Protein Markers in Whole Blood.

Applications

Unspecified application

Species

Unspecified reactive species

Hao Fu,Zhen Qin,Xiao Li,Yueyue Pan,Haitong Xu,Peng Pan,Pengfei Song,Xinyu Liu

Journal of molecular recognition : JMR 36:e2995 PubMed36116102

2022

Electro-immunosensor for ultra-sensitive determination of cardiac troponin I based on reduced graphene oxide and polytyramine.

Applications

Unspecified application

Species

Unspecified reactive species

Jéssica G Brussasco,Pedro H G Guedes,Anna C R Moço,Dayane D Moraes,José M R Flauzino,Heliane S Silva,Atair C Silva,Jonathan P Silva,João M Madurro,Ana G Brito-Madurro

Nature biomedical engineering 6:351-371 PubMed35478225

2022

A multi-organ chip with matured tissue niches linked by vascular flow.

Applications

Unspecified application

Species

Unspecified reactive species

Kacey Ronaldson-Bouchard,Diogo Teles,Keith Yeager,Daniel Naveed Tavakol,Yimu Zhao,Alan Chramiec,Somnath Tagore,Max Summers,Sophia Stylianos,Manuel Tamargo,Busub Marcus Lee,Susan P Halligan,Erbil Hasan Abaci,Zongyou Guo,Joanna Jacków,Alberto Pappalardo,Jerry Shih,Rajesh K Soni,Shivam Sonar,Carrie German,Angela M Christiano,Andrea Califano,Karen K Hirschi,Christopher S Chen,Andrzej Przekwas,Gordana Vunjak-Novakovic

Experimental biology and medicine (Maywood, N.J.) 247:1205-1213 PubMed35466734

2022

Lipocalin-2, S100A8/A9, and cystatin C: Potential predictive biomarkers of cardiovascular complications in COVID-19.

Applications

Unspecified application

Species

Unspecified reactive species

Anamika Gupta,Abaher O Al-Tamimi,Rabih Halwani,Hend Alsaidi,Meganathan Kannan,Firdos Ahmad

Analytical and bioanalytical chemistry 414:5319-5327 PubMed34595559

2021

Electrochemical and electronic detection of biomarkers in serum: a systematic comparison using aptamer-functionalized surfaces.

Applications

Unspecified application

Species

Unspecified reactive species

Vladyslav Mishyn,Teresa Rodrigues,Yann R Leroux,Laura Butruille,Eloise Woitrain,David Montaigne,Patrik Aspermair,Henri Happy,Wolfgang Knoll,Rabah Boukherroub,Sabine Szunerits

Journal of cardiovascular translational research 15:381-390 PubMed34417673

2021

Role of Anti-Beta-1-Adrenergic Receptor Antibodies in Cardiac Dysfunction in Patients with Cirrhotic Cardiomyopathy.

Applications

Unspecified application

Species

Unspecified reactive species

Lixia Ma,Xiaohui Liu,Qingshan Wu,Xing Hu,Hongqun Liu,Jing Zhang,Samuel S Lee

Biomolecules 10: PubMed32784904

2020

Targeting TLR4 with ApTOLL Improves Heart Function in Response to Coronary Ischemia Reperfusion in Pigs Undergoing Acute Myocardial Infarction.

Applications

Unspecified application

Species

Unspecified reactive species

Rafael Ramirez-Carracedo,Laura Tesoro,Ignacio Hernandez,Javier Diez-Mata,David Piñeiro,Macarena Hernandez-Jimenez,Jose Luis Zamorano,Carlos Zaragoza

Nanomaterials (Basel, Switzerland) 10: PubMed32708486

2020

Troponin Aptamer on an Atomically Flat Au Nanoplate Platform for Detection of Cardiac Troponin I.

Applications

Unspecified application

Species

Unspecified reactive species

Hyoban Lee,Hyungjun Youn,Ahreum Hwang,Hyunsoo Lee,Jeong Young Park,Weon Kim,Youngdong Yoo,Changill Ban,Taejoon Kang,Bongsoo Kim
View all publications
websiteProtocolBooklet
en

Product promise

We are committed to supporting your work with high-quality reagents, and we're here for you every step of the way. In the unlikely event that one of our products does not perform as expected, you're protected by our Product Promise.
For full details, please see our Terms & Conditions

Please note: All products are 'FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC OR THERAPEUTIC PROCEDURES'.

For licensing inquiries, please contact partnerships@abcam.com