Human D-Dimer ELISA Kit
- SimpleStep
- What is this?
Be the first to review this product! Submit a review
|
(1 Publication)
- Colorimetric Sandwich ELISA - 450 nm readout : works on any standard plate reader
- Different formats for different needs: 384-well plate for higher throughput
- ELISA
Supplier Data
ELISA - Human D-Dimer ELISA Kit (AB315310)
- sELISA
Supplier Data
Sandwich ELISA - Human D-Dimer ELISA Kit (AB315310)
Interpolated concentrations of D-Dimer in normal human serum donors
Serum of eight individual healthy human female donors was measured in duplicate. Interpolated dilution factor corrected values are plotted (mean +/- SD, n=2). The mean D-Dimer concentration was determined to be 657 ng/mL with a range of 68 – 1,356 ng/mL.
- sELISA
Supplier Data
Sandwich ELISA - Human D-Dimer ELISA Kit (AB315310)
Interpolated concentrations of human D-Dimer in human serum, plasma (citrate), plasma (EDTA), and plasma (Heparin)
Interpolated concentration of native D-Dimer was measured in duplicate at different sample concentrations. Undiluted samples are as follows : Human serum 1 : 200, plasma (citrate) 1 : 1,250, plasma (EDTA) 1 : 25, and plasma (heparin) 1 : 25. The interpolated dilution factor corrected values (to neat sample) are plotted (mean +/- SD, n=2). Sample dilutions are made in Sample Diluent NS.
- sELISA
Supplier Data
Sandwich ELISA - Human D-Dimer ELISA Kit (AB315310)
Interpolated concentrations of human D-Dimer in serum in 96-well vs. 384-well plates
Interpolated concentration of native D-Dimer was measured in duplicate at different sample concentrations in 96-well vs. 384-well plates. Undiluted samples are 0.5% serum. The interpolated dilution factor corrected values are plotted (mean +/- SD, n=2). Sample dilutions are made in Sample Diluent NS.
- sELISA
Supplier Data
Sandwich ELISA - Human D-Dimer ELISA Kit (AB315310)
Example of human D-Dimer standard curve in Sample Diluent NS. Background-subtracted data values (mean +/- SD) are graphed.
- sELISA
Supplier Data
Sandwich ELISA - Human D-Dimer ELISA Kit (AB315310)
Standard curve comparison between the original Human D-Dimer SimpleStep ELISA (ab260076) and current Human D-Dimer SimpleStep ELISA (ab315310).
- sELISA
Supplier Data
Sandwich ELISA - Human D-Dimer ELISA Kit (AB315310)
Example of human D-Dimer standard curve in Sample Diluent NS in 96-well vs. 384-well plate. Background-subtracted data values (mean +/- SD) are graphed.
Reactivity data
Product details
How the assay works
Human D-Dimer SimpleStep ELISA®employs capture antibodies conjugated to an affinity tag that is recognized by the monoclonal antibody used to coat our SimpleStep ELISA® plates. This approach to sandwich ELISA allows the formation of the antibody-analyte sandwich complex in a single step, significantly reducing assay time. See the SimpleStep ELISA® protocol summary in the image section for further details.
Assay Specificity
Our SimpleStep ELISA® kits use recombinant monoclonal antibodies rigorously validated to ensure the highest level of consistency and reproducibility, improved sensitivity and specificity and ease of scalability and security of supply.
Please refer to our protocol booklet for more details.
Human D-Dimer ELISA Kit ab315310 protocol summary
1. Mix: add samples/standards to the wells together with the capture and detector antibody cocktail. Incubate 1 hr at room temperature
2. Wash
3. Add TMB development solution - incubate for 10 min
4. Add Stop solution
5. Read the results on a plate reader at 450 nm
Save your precious samples
Our 384-well Human D-Dimer ELISA Kit only requires a maximum of 12.5µL sample
Precision & recovery information
Precision
[ { "reproducibilityType": "Intra", "sample": "Serum", "replicates": 8, "mean": null, "standardDeviation": null, "coefficientOfVariability": "= 7.4" }, { "reproducibilityType": "Inter", "sample": "Serum", "replicates": 3, "mean": null, "standardDeviation": null, "coefficientOfVariability": "= 3.8" } ]
Recovery
[ { "sample": "Serum", "range": "92 - 98 %", "average": "= 95" }, { "sample": "EDTA Plasma", "range": "101 - 113 %", "average": "= 105" }, { "sample": "Citrate plasma", "range": "99 - 113 %", "average": "= 106" }, { "sample": "Heparin Plasma", "range": "97 - 104 %", "average": "= 101" }, { "sample": "Urine", "range": "85 - 93 %", "average": "= 90" }, { "sample": "Cell culture media", "range": "86 - 99 %", "average": "= 92" } ]
What's included?
Properties and storage information
Shipped at conditions
Appropriate short-term storage conditions
Appropriate long-term storage conditions
Storage information
Product protocols
- Visit the General protocols
- Visit the Troubleshooting
- Download websiteProtocolBooklet|
Target data
Publications (1)
Recent publications for all applications. Explore the full listand refine your search
Biomedical reports 20:96 PubMed38765860
2024
Applications
Unspecified application
Species
Unspecified reactive species
Product promise
Please note: All products are 'FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC OR THERAPEUTIC PROCEDURES'.
For licensing inquiries, please contact partnerships@abcam.com