Skip to main content

Human Metabotropic Glutamate Receptor 3/MGLUR3 ELISA Kit is a single-wash 90-min Simplestep used to quantify Human Metabotropic Glutamate Receptor 3/MGLUR3 with a sensitivity of 9.14 pg/ml. The assay uses a simple mix-wash-read protocol with just one incubation and one wash step.

- Colorimetric Sandwich ELISA - 450 nm readout : works on any standard plate reader
- Design your own immunoassay: we also offer the conjugation-ready antibody pair

Be the first to review this product! Submit a review

Images

Sandwich ELISA - Human Metabotropic Glutamate Receptor 3/MGLUR3 ELISA Kit (AB309279), expandable thumbnail
  • Sandwich ELISA - Human Metabotropic Glutamate Receptor 3/MGLUR3 ELISA Kit (AB309279), expandable thumbnail

Key facts

Detection method
Colorimetric
Sample types
Tissue Extracts
Assay type
Sandwich (quantitative)
Reactive species
Human
Range
31.25 - 2000 pg/mL
Assay time
1h 30m
Sensitivity
= 9.14 pg/mL

Reactivity data

Application
sELISA
Reactivity
Reacts
Dilution info
-
Notes

-

Target data

Function

G-protein coupled receptor for glutamate. Ligand binding causes a conformation change that triggers signaling via guanine nucleotide-binding proteins (G proteins) and modulates the activity of down-stream effectors. Signaling inhibits adenylate cyclase activity.

Alternative names

What's included?

1 x 96 Tests
Components
10X Human Metabotropic Glutamate Receptor 3/MGLUR3 Capture Antibody
1 x 600 µL
10X Human Metabotropic Glutamate Receptor 3/MGLUR3 Detector Antibody
1 x 600 µL
10X Wash Buffer PT (ab206977)
1 x 20 mL
5X Cell Extraction Buffer PTR (ab193970)
1 x 10 mL
Antibody Diluent 4BR
1 x 6 mL
Human Metabotropic Glutamate Receptor 3/MGLUR3 Lyophilized Recombinant Protein
2 x 1 Vial
Plate Seals
1 x 1 Unit
Sample Diluent NS (ab193972)
1 x 12 mL
SimpleStep Pre-Coated 96-Well Microplate (ab206978)
1 x 1 Unit
Stop Solution
1 x 12 mL
TMB Development Solution
1 x 12 mL

Recommended products

Human Metabotropic Glutamate Receptor 3/MGLUR3 ELISA Kit is a single-wash 90-min Simplestep used to quantify Human Metabotropic Glutamate Receptor 3/MGLUR3 with a sensitivity of 9.14 pg/ml. The assay uses a simple mix-wash-read protocol with just one incubation and one wash step.

- Colorimetric Sandwich ELISA - 450 nm readout : works on any standard plate reader
- Design your own immunoassay: we also offer the conjugation-ready antibody pair

Key facts

Detection method
Colorimetric
Sample types
Tissue Extracts
Assay type
Sandwich (quantitative)
Reactive species
Human
Range
31.25 - 2000 pg/mL
Assay time
1h 30m
Assay Platform
Pre-coated microplate (12 x 8 well strips)
Sensitivity
= 9.14 pg/mL
General recovery
103%

Precision

Intra assay

Sample
Extract
n
8
mean
-
SD
-
C.V.
10.3

Inter assay

Sample
Extract
n
3
mean
-
SD
-
C.V.
5.3

Recovery

Sample specific recovery

Sample type
Tissue Extracts
Average %
= 103
Range
97 - 113 %

Storage

Shipped at conditions
Blue Ice
Appropriate short-term storage conditions
+4°C
Appropriate long-term storage conditions
+4°C
Storage information
+4°C

Notes

Human Metabotropic Glutamate Receptor 3/MGLUR3 SimpleStep ELISA® kit is a single-wash 90 min sandwich ELISA designed for the quantitative measurement of Metabotropic Glutamate Receptor 3/MGLUR3 protein in human serum, plasma, saliva and cell culture supernatant. Quantitate Human Metabotropic Glutamate Receptor 3/MGLUR3 with 9.14 pg/ml sensitivity.

SimpleStep ELISA® technology employs capture antibodies conjugated to an affinity tag that is recognized by the monoclonal antibody used to coat our SimpleStep ELISA® plates. This approach to sandwich ELISA allows the formation of the antibody-analyte sandwich complex in a single step, significantly reducing assay time. See the SimpleStep ELISA® protocol summary in the image section for further details. Our SimpleStep ELISA® technology provides several benefits:

-Single-wash protocol reduces assay time to 90 minutes or less
-High sensitivity, specificity and reproducibility from superior antibodies
-Fully validated in biological samples
-96-wells plate breakable into 12 x 8 wells strips

A 384-well SimpleStep ELISA® microplate (Pre-coated 384 well Microplate SimpleStep ELISA® ab203359) is available to use as an alternative to the 96-well microplate provided with SimpleStep ELISA® kits.

Abcam has not and does not intend to apply for the REACH Authorisation of customers' uses of products that contain European Authorisation list (Annex XIV) substances.
It is the responsibility of our customers to check the necessity of application of REACH Authorisation, and any other relevant authorisations, for their intended uses.

Product promise

We are dedicated to supporting your work with high quality reagents and we are here for you every step of the way should you need us.

In the unlikely event of one of our products not working as expected, you are covered by our product promise.

Full details and terms and conditions can be found here:
Terms & Conditions.

2 product images

  • Sandwich ELISA - Human Metabotropic Glutamate Receptor 3/MGLUR3 ELISA Kit (ab309279), expandable thumbnail

    Sandwich ELISA - Human Metabotropic Glutamate Receptor 3/MGLUR3 ELISA Kit (ab309279)

    Example of human Metabotropic Glutamate Receptor 3/MGLUR3 standard curve. Background-subtracted data values (mean +/- SD) are graphed. 

  • Sandwich ELISA - Human Metabotropic Glutamate Receptor 3/MGLUR3 ELISA Kit (ab309279), expandable thumbnail

    Sandwich ELISA - Human Metabotropic Glutamate Receptor 3/MGLUR3 ELISA Kit (ab309279)

    Interpolated concentration of native Metabotropic Glutamate Receptor 3/MGLUR3 was measured in duplicate at different sample concentrations. Undiluted samples are as follows: human hippocampus extract 200 μg/mL, mouse brain extract 100 μg/mL, and rat brain extract 200 μg/mL. The interpolated dilution factor corrected values are plotted (mean +/- SD, n=2). Sample dilutions are made in 1X Cell Extraction Buffer PTR.

Downloads

Product protocols

For this product, it's our understanding that no specific protocols are required. You can:

Please note: All products are 'FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC OR THERAPEUTIC PROCEDURES'.

For licensing inquiries, please contact partnerships@abcam.com