AB326984

Extracellular Vesicle (Ev) SEC Columns 70 nm

Be the first to review this product! Submit a review

|

(0 Publication)

Extracellular Vesicle (Ev) SEC Columns 70 nm (ab326984) is intended to be used for EVs isolation from biological fluids (Plasma, Serum, Urine, CSF, Cell Culture Media).
1 Images
Other - Extracellular Vesicle (Ev) SEC Columns 70 nm (AB326984)
  • Other

Supplier Data

Other - Extracellular Vesicle (Ev) SEC Columns 70 nm (AB326984)

A comparison was made using ab326984 and a competitor, obtaining a better performance in terms of recovery.
Top image : Comparison of the diameter of the isolated particles by column type. The D50 value means 50% of the particles have that diameter or less.

Bottom image : Concentration (particles / ml) obtained from purifying plasma (500 µl).

Key facts

Form

Solid

form

Storage buffer

Preservative: 0.1% Sodium azide

storage-buffer

Product details

It is recommended to use buffers with an ionic strength of 0.15 M or higher, because in low ionic strength buffers, the concentration of competing buffer ions is low, and contaminant proteins spend most of their time adsorbed to binding sites on the stationary phase. For buffer choice it is also important consider the downstream protocol for isolated vesicles. For example citrate buffers could interfere with molecular biology protocols.

Note: All buffers must contain sodium azide.

Columns are reusable up to 5 times and should not be used if any evidence of deterioration is observed.

PROTOCOL
1. Place the column in a rack and level it, make sure the column is vertical.
2. Remove the top cap carefully, as drastic changes in pressure could damage the column.
3. Remove the bottom cap.

Column Equilibration
4. The column must be at room temperature (18ºC – 24ºC) for 30 minutes before use.
5. Use only fresh filtered (0.2 μm) buffer to avoid particulate contamination.
6. Rinse the column with 21 mL of running buffer (7,5 mL for the lower volumes columns).
7. Do not allow the column to run dry. The top filter must stay wet.

Sample fraction collection
8. Once the column is washed, add the sample onto the top filter (500 μL, and 150 μL for the lower volumes columns). Once it passes through the filter, fill the column with the running buffer.
9. Let pass three drops of running buffer, and collect fractions of 500 μL (150 μL for the lower volumes columns) (minimum 12 fractions).
10. When the column is used according to the protocol, the first four fractions (2.0 mL, and 60 0 μL for the lower volumes columns) is the void volume which does not contain vesicles. These vesicles elute predominantly in fractions 5, 6, 7 and 8 with removal of protein contamination. Fractions beyond 9 usually contain higher protein and lower vesicle levels.

COLUMN CLEANING AND STORAGE
1. After the collection of the vesicle fraction, rinse the column with 21 mL (7,5 mL for the lower volumes columns) of running buffer to rinse out all the protein and small molecules before the next sample application.
2. NaOH cleanup is performed to remove precipitated proteins or other contaminants that may be accumulated on the column, a running buffer change, or a sample change. To do this, add 0.5mL (0,15 mL for the lower volumes columns) of sodium hydroxide (500 mM NaOH), then equilibrate the columns with 10 mL of water (3,5 mL for the lower volumes columns) followed by 21 mL of running buffer (7,5 mL for the lower volumes columns) immediately after cleanup.
Note that the column should never be stored in sodium hydroxide.
3. Put the bottom cap on.
4. Put the top cap on carefully, as drastic changes in pressure could damage the column.
5. Stopcocks cannot be heat sterilized as they do not withstand high temperatures. To sterilize them, it can be done by passing a 30% ethanol solution (5 mL) with the help of a syringe.

FRACTION COLLECTION CONCENTRATION
If the volume of the starting sample is small and / or the concentration and EVs is low in the sample, it is possible that the EVs will be much diluted in the collected fractions. In many cases, for downstream processing it is necessary to concentrate the fractions.

Properties and storage information

Shipped at conditions
Ambient - Can Ship with Ice
Appropriate short-term storage conditions
+4°C
Appropriate long-term storage conditions
+4°C
Storage information
Store in the dark

Product protocols

Target data

Product promise

We are committed to supporting your work with high-quality reagents, and we're here for you every step of the way. In the unlikely event that one of our products does not perform as expected, you're protected by our Product Promise.
For full details, please see our Terms & Conditions

Please note: All products are 'FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC OR THERAPEUTIC PROCEDURES'.

For licensing inquiries, please contact partnerships@abcam.com