Designing a multicolor flow cytometry protocol

Increasing the number of antigens and fluorophores can also increase the complexity of your experiment.

While flow cytometry is a powerful tool for identifying and analyzing multiple antigens simultaneously, increasing the number of antigens and fluorophores also increases the complexity of the experimental design. Once you know which antigens you want to study, you need to build your panel. Usually, this is the most challenging aspect of setting up a multicolor flow cytometry experiment. The steps below aim to provide a quick and easy guide to help you in the process.1

Flow cytometry application guide

PDF

Download
button-secondary
icon-none

Know your flow cytometer

Before designing your multicolor flow panel, you will need to determine the following factors:

  1. The number and type of lasers
  2. The number of detectors
  3. The type of filters available on your flow cytometer

You will only be able to use fluorophores excited by the corresponding wavelength of light from the laser (Figure 13).

Figure 13. Common lasers used in flow cytometry

Figure 13. Common lasers used in flow cytometry

You should match the emission wavelength of your fluorophores to the available filters and the wavelengths they allow to pass. You can check the relative brightness of fluorophores and their excitation and emission spectra using our fluorophore chart.

Refer to your instrument’s manual or speak to your core facility manager to ensure optimal detection.

Know your cell populations, antigens, and fluorophores

Some cell populations are rare, or antigen density is low due to functional differences and cell activation levels.

A highly expressed antigen will usually be detected and resolved from the negative control with almost any fluorophore. Generally, when designing a multicolor flow cytometry panel, use the brightest fluorophores (such as PE or APC) for low or unknown antigen expression targets or rare cell populations (Figure 14) and dimmer fluorophores to detect higher abundance targets.

Figure 14. An antigen expressed at lower density might require a higher signal to background ratio provided by a brighter PE or APC conjugate to separate positive cells adequately from unstained cells.

Figure 14. An antigen expressed at lower density might require a higher signal to background ratio provided by a brighter PE or APC conjugate to separate positive cells adequately from unstained cells.

Bear in mind that other factors, such as sample buffer pH, can affect fluorophore brightness. The relative fluorophore intensity also depends on the instrument due to differences in the laser and filter combinations. Be sure to use the appropriate FACS instrument for the fluorophores you wish to detect.

Minimize spectral overlap

For best results with multicolor flow cytometry, choose the brightest fluorophores with little or no overlap between their emission spectra (Figure 15). You can use compensation to control the effects of spectral overlap (see below), but it is worth sacrificing some brightness in one detector to avoid spillover (Table 3).

Figure 15. Where possible, choose fluorophores with little or no overlap in their emission spectra.

Figure 15. Where possible, choose fluorophores with little or no overlap in their emission spectra.

Table 3. Examples of good and poor fluorochrome combinations for flow cytometry.

Fluorochrome
Target Expression
Lasers
Channels
Brightness
Compensation
Combination
FITC
APC
High
Low
Blue
Red
Green
Red
Medium
High
Mild
Good
FITC
PE
High
Low
Blue
Yellow
Green
Yellow
Medium
High
Moderate
Medium
PerCP
7–AAD
High
Low
Blue
Blue
Red
Red
Low
Medium
Moderate
Poor

Fluorescence compensation

Fluorophores often have some overlap in their emission spectra, as demonstrated in the example below for PE and FITC. This overlap can generate a false positive signal as fluorescence from more than one fluorophore may be detected in a single channel. Spectral overlap can be problematic in multicolor experiments. Therefore, it must be corrected using compensation to ensure that the fluorescence detected is genuinely derived from the measured fluorophore.2

Before compensation

Following excitation with the blue laser (488 nm), FITC emission is primarily detected in the channel-specific for FITC, but the emission tail lies within the range of the filter used to detect PE. These are seen as false positive signals in the PE channel, meaning that cells positive for FITC will appear positive for PE (Figure 16).

Figure 16. False-positive PE signals generated by FITC due to spectral overlap

Figure 16. False-positive PE signals generated by FITC due to spectral overlap

After compensation

To compensate, a sample stained only with a FITC-labeled antibody is required. The settings can then be adjusted until no FITC signal is seen in the PE channel. Following compensation, FITC emission is solely detected in the channel specific for FITC (Figure 17).

Figure 17. Compensation removes false positive signals caused by spectral overlap.

Figure 17. Compensation removes false positive signals caused by spectral overlap.

Hints and tips for applying compensation in multicolor flow cytometry

The procedure for setting the correct fluorescence compensation follows the same basic principles on any cytometer. However, due to subtle differences between instruments, we always recommend reviewing the manufacturer’s instructions. These are some valuable principles to follow:

Compensation controls are required for each fluorophore and should contain a positive and negative population. The following guidelines should be applied:

References

  1. Busuttil Crellin, X.,, McCafferty, C.,, Van Den Helm, S.,, et al. Guidelines for panel design, optimization, and performance of whole blood multicolor flow cytometry of platelet surface markers Platelets  31 (7),845-852 (2020)
  2. Szaloki, G.,, Goda, K. Compensation in multicolor flow cytometry Cytometry A 87 (11),982-985 (2015)