Getting the most out of your SimpleStep Ignite™ ELISA kit

Our best practice tips to help you get the best sensitivity and precision out of your SimpleStep Ignite™ ELISA kit

Introduction

SimpleStep Ignite™ assays are designed to deliver excellent sensitivity, precision, and ease of use. While the standard protocol provides robust performance for most applications, some customers require exceptionally low limits of detection and exceptionally tight precision requirements. Through collaborations with users working at the limits of assay performance, we have identified several best practices that can help maximize sensitivity, reduce variability, and improve overall reproducibility.

Table of contents

Washing strategy

Incubation time

Pipetting technique

Additional considerations

Summary

Optimize your washing strategy

Effective washing is one of the most important factors influencing assay performance, particularly when measuring very low analyte concentrations. Customers seeking the lowest background, best signal-to-noise ratios, and tightest precision have often benefited from careful attention to the following washing parameters and practices:

Even modest increases in washing stringency can reduce nonspecific binding, lower background signal, improve low-end sensitivity, and enhance replicate precision.

To maximize wash efficiency:

Additional considerations for manual washing

Sufficient wash buffer volume is critical for optimal SimpleStep Ignite™ ELISA performance. Standard curves were generated using wash volumes of 200, 250, 300, and 350 µL per well per wash step. Reducing wash volume below the recommended 350 µL per well increased background, reduced signal-to-noise, and resulted in greater variability, particularly near the lower end of the assay range. For optimal sensitivity, precision, and assay robustness, use 350 µL per well per wash step and do not reduce the recommended wash volume due to concerns about minor overflow. Under the recommended wash conditions, minor overflow is unlikely to meaningfully affect assay performance; however, excessive overflow or splashing between wells should be avoided.

For customers performing manual wash steps, effective removal of residual liquid between washes is  especially important.

After each decanting step, gently invert the plate and tap it onto a clean, absorbent laboratory wipe to remove any remaining wash buffer. Residual liquid left in wells can dilute subsequently added reagents, contribute to increased background, and introduce well-to-well variability. Care should be taken to avoid excessive force during tapping, as this may cause splashing or cross-contamination between wells, leading to increased well-to-well variability and higher CVs.

For 96-well strip plates, label individual strips before starting the assay and periodically verify that all strips remain fully seated in the plate frame throughout the washing process.

Considerations for increased wash cycles

Customers pursuing maximum sensitivity and precision often choose to increase the number of wash cycles beyond the standard protocol. When performing additional washes, the supplied wash buffer may become a limiting reagent.

In these situations, 10X Wash Buffer PT (ab206977) may be used to support increased wash volumes. This can be particularly advantageous for customers using automated plate washers, where bulk wash solutions simplify workflow and reduce the need for frequent reagent replacement. This buffer may also be a practical option when performing additional manual wash steps as part of assay optimization.

Regardless of the wash solution used, consistent execution of wash steps remains critical for achieving optimal assay performance. Adequate soak times, complete aspiration, effective residual liquid removal, and sufficient wash volumes all contribute to lower background, improved sensitivity, and better assay precision.

Read time window after substrate addition

SimpleStep Ignite™ ELISA assays are designed to provide a stable chemiluminescent signal after substrate addition, allowing flexibility in plate handling and reading.

Best practice: Read the microplate within 15 minutes of substrate addition. Immediate reading is preferred, but comparable performance has been observed up to 15 minutes after substrate addition.

Validation studies across multiple Ignite assays demonstrated:

Optimize incubation time

For certain low-abundance targets, particularly when analyzing complex, protein-rich sample matrices such as serum or plasma, extending the incubation time can further improve assay performance.

While the standard protocol is optimized to balance performance and workflow efficiency, some customers have observed improvements in low-end precision and signal recovery by increasing the incubation time up to 2 hours. The extended incubation period provides additional opportunity for target binding, which can be particularly beneficial when detecting analytes near the lower end of the assay range or when matrix components may slow binding interactions.

Potential benefits of extended incubation include:

Because the impact of longer incubations can vary by target and sample type, we recommend evaluating this modification during assay optimization to determine whether it provides measurable benefits for your specific application.

Example data

Optional incubation extension for performance optimization. Representative IL-2 SimpleStep Ignite™ ELISA data showing the effect of increasing incubation time from 15 to 60 minutes. While the standard protocol provides the validated assay performance specifications, additional incubation time may further improve signal intensity, signal-to-noise ratio, and low-end sensitivity for some assays. The extent of improvement is target and matrix dependent.

In this example, extending incubation time from 15 to 60 minutes improved the LOD from approximately 4.8 pg/mL to 3.0 pg/mL, corresponding to an approximately 1.6-fold improvement in analytical sensitivity. The longer incubation also improved signal-to-noise ratios across low, medium, and high analyte concentrations, providing better separation between blank and low-concentration samples.

The magnitude of improvement can vary by target and sample matrix. For some assays, extending incubation time may yield only modest gains, while others may show substantially larger improvements, with sensitivity improvements of up to 5-fold observed in certain applications.

Most of the benefit is typically achieved within the first 30-60 minutes of additional incubation, while gains beyond 2 hours are often limited. A 30-minute incubation provides a strong balance between assay performance and workflow efficiency for many applications.

Optimize pipetting technique

When evaluating assay variability, pipetting technique is often the single largest contributor to elevated CVs. Even highly optimized assays can exhibit increased variability if sample handling practices are not tightly controlled.

Best practices for mimizing CVs

For the most consistent and reproducible results:

Tip changes matter

Of all pipetting variables, changing tips between each transfer has consistently provided the greatest improvement in assay precision.

Fresh tips help minimize:

While tip conservation strategies may be appropriate in some workflows, customers pursuing the tightest CVs and highest reproducibility should use a fresh tip for every sample and reagent addition step.

Additional considerations

To further improve assay performance:

Key recommendations for maximum performance

For customers requiring the highest levels of sensitivity and precision from their SimpleStep Ignite™ assays, we recommend focusing on three key areas:

1. Enhance washing stringency

2. Optimize incubation conditions

3. Follow rigorous pipetting practices