Getting the most out of your SimpleStep Ignite™ ELISA kit
Our best practice tips to help you get the best sensitivity and precision out of your SimpleStep Ignite™ ELISA kit
Introduction
SimpleStep Ignite™ assays are designed to deliver excellent sensitivity, precision, and ease of use. While the standard protocol provides robust performance for most applications, some customers require exceptionally low limits of detection and exceptionally tight precision requirements. Through collaborations with users working at the limits of assay performance, we have identified several best practices that can help maximize sensitivity, reduce variability, and improve overall reproducibility.
Table of contents
Optimize your washing strategy
Effective washing is one of the most important factors influencing assay performance, particularly when measuring very low analyte concentrations. Customers seeking the lowest background, best signal-to-noise ratios, and tightest precision have often benefited from careful attention to the following washing parameters and practices:
- Increasing wash, soak or wait times between dispense and aspiration steps
- Plate shaking during soak times at the recommended RPM to dislodge bubbles
- Adding additional wash cycles beyond the standard protocol
- Ensuring complete aspiration and removal of residual liquid between washes
- Verifying that automated plate washers are properly aligned and calibrated
Even modest increases in washing stringency can reduce nonspecific binding, lower background signal, improve low-end sensitivity, and enhance replicate precision.
To maximize wash efficiency:
- Use the recommended wash volume (350 μL) for every wash cycle. Do not reduce wash volumes to prevent minor overflow at the top of wells. Complete well filling is essential for effective washing, and lower wash volumes may increase background, variability, and reduce assay sensitivity.
- Allow adequate soak time between dispense and aspiration steps. Optionally, shaking the plate between aspiration and dispense steps may provide a more complete wash step. Allow one to two minutes soak/shake time before aspiration.
- Ensure complete aspiration before proceeding to the next wash cycle or assay step. Proceed promptly to the next assay step after aspiration or blotting. Do not allow wells to dry between assay steps, as drying can adversely affect assay performance and may result in increased variability, elevated background, reduced signal recovery, and worsen overall precision.
Additional considerations for manual washing
Sufficient wash buffer volume is critical for optimal SimpleStep Ignite™ ELISA performance. Standard curves were generated using wash volumes of 200, 250, 300, and 350 µL per well per wash step. Reducing wash volume below the recommended 350 µL per well increased background, reduced signal-to-noise, and resulted in greater variability, particularly near the lower end of the assay range. For optimal sensitivity, precision, and assay robustness, use 350 µL per well per wash step and do not reduce the recommended wash volume due to concerns about minor overflow. Under the recommended wash conditions, minor overflow is unlikely to meaningfully affect assay performance; however, excessive overflow or splashing between wells should be avoided.
For customers performing manual wash steps, effective removal of residual liquid between washes is especially important.
After each decanting step, gently invert the plate and tap it onto a clean, absorbent laboratory wipe to remove any remaining wash buffer. Residual liquid left in wells can dilute subsequently added reagents, contribute to increased background, and introduce well-to-well variability. Care should be taken to avoid excessive force during tapping, as this may cause splashing or cross-contamination between wells, leading to increased well-to-well variability and higher CVs.
For 96-well strip plates, label individual strips before starting the assay and periodically verify that all strips remain fully seated in the plate frame throughout the washing process.
Considerations for increased wash cycles
Customers pursuing maximum sensitivity and precision often choose to increase the number of wash cycles beyond the standard protocol. When performing additional washes, the supplied wash buffer may become a limiting reagent.
In these situations, 10X Wash Buffer PT (ab206977) may be used to support increased wash volumes. This can be particularly advantageous for customers using automated plate washers, where bulk wash solutions simplify workflow and reduce the need for frequent reagent replacement. This buffer may also be a practical option when performing additional manual wash steps as part of assay optimization.
Regardless of the wash solution used, consistent execution of wash steps remains critical for achieving optimal assay performance. Adequate soak times, complete aspiration, effective residual liquid removal, and sufficient wash volumes all contribute to lower background, improved sensitivity, and better assay precision.
Read time window after substrate addition
SimpleStep Ignite™ ELISA assays are designed to provide a stable chemiluminescent signal after substrate addition, allowing flexibility in plate handling and reading.
Best practice: Read the microplate within 15 minutes of substrate addition. Immediate reading is preferred, but comparable performance has been observed up to 15 minutes after substrate addition.
Validation studies across multiple Ignite assays demonstrated:
- Standard curves generated immediately after substrate addition and up to 15 minutes later showed highly comparable curve shapes and assay performance.
- While some assays remain usable beyond 15 minutes, extended delays (eg 30 minutes or longer) may alter curve shape and reduce quantitative accuracy. Changes are most commonly observed at the upper end of the standard curve but may also occur at lower concentrations depending on the assay.
- Assay-specific performance may vary, therefore minimizing the delay between substrate addition and plate reading is recommended for optimal quantitative accuracy.
Optimize incubation time
For certain low-abundance targets, particularly when analyzing complex, protein-rich sample matrices such as serum or plasma, extending the incubation time can further improve assay performance.
While the standard protocol is optimized to balance performance and workflow efficiency, some customers have observed improvements in low-end precision and signal recovery by increasing the incubation time up to 2 hours. The extended incubation period provides additional opportunity for target binding, which can be particularly beneficial when detecting analytes near the lower end of the assay range or when matrix components may slow binding interactions.
Potential benefits of extended incubation include:
- Improved low-end sensitivity
- Better separation between blank and low-concentration samples
- Reduced variability among low-level replicates
- Improved recovery in protein-dense matrices
Because the impact of longer incubations can vary by target and sample type, we recommend evaluating this modification during assay optimization to determine whether it provides measurable benefits for your specific application.
Example data
Optional incubation extension for performance optimization. Representative IL-2 SimpleStep Ignite™ ELISA data showing the effect of increasing incubation time from 15 to 60 minutes. While the standard protocol provides the validated assay performance specifications, additional incubation time may further improve signal intensity, signal-to-noise ratio, and low-end sensitivity for some assays. The extent of improvement is target and matrix dependent.
In this example, extending incubation time from 15 to 60 minutes improved the LOD from approximately 4.8 pg/mL to 3.0 pg/mL, corresponding to an approximately 1.6-fold improvement in analytical sensitivity. The longer incubation also improved signal-to-noise ratios across low, medium, and high analyte concentrations, providing better separation between blank and low-concentration samples.
The magnitude of improvement can vary by target and sample matrix. For some assays, extending incubation time may yield only modest gains, while others may show substantially larger improvements, with sensitivity improvements of up to 5-fold observed in certain applications.
Most of the benefit is typically achieved within the first 30-60 minutes of additional incubation, while gains beyond 2 hours are often limited. A 30-minute incubation provides a strong balance between assay performance and workflow efficiency for many applications.
Optimize pipetting technique
When evaluating assay variability, pipetting technique is often the single largest contributor to elevated CVs. Even highly optimized assays can exhibit increased variability if sample handling practices are not tightly controlled.
Best practices for mimizing CVs
For the most consistent and reproducible results:
- Change pipette tips between every sample and reagent transfer
- Use calibrated pipettes and maintain them regularly
- Pre-wet tips when handling small volumes
- Pipette slowly and consistently to avoid bubble formation
- Ensure reagents and samples are thoroughly mixed before dispensing
- Use a consistent pipetting angle and technique across the entire plate
- Avoid touching the sides of wells when dispensing samples or reagents
Tip changes matter
Of all pipetting variables, changing tips between each transfer has consistently provided the greatest improvement in assay precision.
Fresh tips help minimize:
- Carryover contamination
- Residual reagent effects
- Inconsistent delivered volumes
- Cross-contamination between standards, controls, and samples
While tip conservation strategies may be appropriate in some workflows, customers pursuing the tightest CVs and highest reproducibility should use a fresh tip for every sample and reagent addition step.
Additional considerations
To further improve assay performance:
- Equilibrate reagents to recommended operating temperatures before use
- Follow incubation times consistently across all wells
- Avoid plate drying during assay execution
- Maintain a clean workspace to minimize contamination risks
- Validate automated liquid handling methods before high-volume implementation
Key recommendations for maximum performance
For customers requiring the highest levels of sensitivity and precision from their SimpleStep Ignite™ assays, we recommend focusing on three key areas:
1. Enhance washing stringency
- Increase wash soak/wait times.
- Add additional wash cycles as needed.
- Ensure complete aspiration between washes.
- Consider using bulk PBST when higher wash volumes are desired.
2. Optimize incubation conditions
- Evaluate extending incubation times up to 2 hours, particularly for low-abundance targets in complex biological matrices such as serum or plasma.
- Assess the impact on low-end precision and analyte recovery during assay optimization.
3. Follow rigorous pipetting practices
- Use calibrated pipettes and maintain consistent technique.
- Thoroughly mix samples and reagents before dispensing.
- Change tips between every sample and reagent transfer to minimize carryover and maximize reproducibility.