Overview

  • Product name
    Human Albumin ELISA Kit
    See all Albumin kits
  • Detection method
    Colorimetric
  • Precision
    Intra-assay
    Sample n Mean SD CV%
    Serum 9 4%
    Inter-assay
    Sample n Mean SD CV%
    Serum 3 12.6%
  • Sample type
    Urine, Serum, Heparin Plasma, EDTA Plasma, Citrate Plasma
  • Assay type
    Sandwich (quantitative)
  • Sensitivity
    30 pg/ml
  • Range
    312.5 pg/ml - 20000 pg/ml
  • Recovery

    Sample specific recovery
    Sample type Average % Range
    Urine 104 101% - 107%
    Serum 117 106% - 133%
    Heparin Plasma 111 104% - 116%
    EDTA Plasma 115 108% - 121%
    Citrate Plasma 112 106% - 117%

  • Assay time
    1h 30m
  • Assay duration
    One step assay
  • Species reactivity
    Reacts with: Human
    Does not react with: Mouse, Rat, Cow, Rhesus monkey
  • Product overview

    Albumin in vitro SimpleStep ELISA® (Enzyme-Linked Immunosorbent Assay) kit is designed for the quantitative measurement of Albumin levels in human plasma, serum, and urine.


    The SimpleStep ELISA® employs an affinity tag labeled capture antibody and a reporter conjugated detector antibody which immunocapture the sample analyte in solution. This entire complex (capture antibody/analyte/detector antibody) is in turn immobilized via immunoaffinity of an anti-tag antibody coating the well. To perform the assay, samples or standards are added to the wells, followed by the antibody mix. After incubation, the wells are washed to remove unbound material. TMB substrate is added and during incubation is catalyzed by HRP, generating blue coloration. This reaction is then stopped by addition of Stop Solution completing any color change from blue to yellow. Signal is generated proportionally to the amount of bound analyte and the intensity is measured at 450 nm. Optionally, instead of the endpoint reading, development of TMB can be recorded kinetically at 600 nm.


    Albumin (ALB), the main protein of plasma, has a good binding capacity for water, Ca2+, Na+, K+, fatty acids, hormones, bilirubin and drugs. Its main function is the regulation of the colloidal osmotic pressure of blood. It also serves as a major zinc transporter in plasma, typically binding about 80% of all plasma zinc.

  • Tested applications
    Suitable for: Sandwich ELISAmore details
  • Platform
    Pre-coated microplate (12 x 8 well strips)

Properties

Applications

Our Abpromise guarantee covers the use of ab227933 in the following tested applications.

The application notes include recommended starting dilutions; optimal dilutions/concentrations should be determined by the end user.

Application Abreviews Notes
Sandwich ELISA Use at an assay dependent concentration.

Images

  • Background-subtracted data values (mean +/- SD) are graphed.

  • The concentrations of Serum Albumin were measured in duplicates, interpolated from the Serum Albumin standard curves and corrected for sample dilution. Undiluted samples are as follows: serum 1:5 million, plasma (citrate) 1:5 million, plasma (heparin) 1:5 million, and plasma (EDTA) 1:10 million. The interpolated dilution factor corrected values are plotted (mean +/- SD, n=2). The mean Serum Albumin concentration was determined to be 39.3 mg/mL in serum, 34.4 mg/mL in plasma (citrate), 30.8 mg/mL in plasma (heparin), and 34.5 mg/mL in plasma (EDTA).

  • The concentrations of Serum Albumin were measured in duplicates, interpolated from the Serum Albumin standard curves and corrected for sample dilution. Undiluted samples are 1:500. The interpolated dilution factor corrected values are plotted (mean +/- SD, n=2). The mean Serum Albumin concentration was determined to be 6,290 ng/mL in urine.

  • Interpolated dilution factor corrected values are plotted (mean +/- SD, n=2). The mean Serum Albumin concentration was determined to be 33.31 mg/mL with a range of 22.5 – 46.2 mg/mL.

Protocols

References

ab227933 has not yet been referenced specifically in any publications.