Human Cardiac Troponin T ELISA Kit is a single-wash 90-min Simplestep used to quantify Human Cardiac Troponin T with a sensitivity of 68.2 pg/ml. The assay uses a simple mix-wash-read protocol with just one incubation and one wash step.

- Colorimetric Sandwich ELISA - 450 nm readout : works on any standard plate reader
- Design your own immunoassay: we also offer the conjugation-ready antibody pair
3 이미지
ELISA - Human Cardiac Troponin T ELISA Kit (AB223860)
  • ELISA

Supplier Data

ELISA - Human Cardiac Troponin T ELISA Kit (AB223860)
Sandwich ELISA - Human Cardiac Troponin T ELISA Kit (AB223860)
  • sELISA

Supplier Data

Sandwich ELISA - Human Cardiac Troponin T ELISA Kit (AB223860)

Interpolated concentrations of native Cardiac Troponin T in human heart tissue extract based on a 100 μg/mL extract load.

The concentrations of Cardiac Troponin T were measured in duplicate and interpolated from the Cardiac Troponin T standard curve and corrected for sample dilution. The interpolated dilution factor corrected values are plotted (mean +/- SD, n=2). The mean Cardiac Troponin T concentration was determined to be 378.8 pg/mL in heart tissue extract

Sandwich ELISA - Human Cardiac Troponin T ELISA Kit (AB223860)
  • sELISA

Supplier Data

Sandwich ELISA - Human Cardiac Troponin T ELISA Kit (AB223860)

Example of human Cardiac Troponin T standard curve in Sample Diluent NS.

The Cardiac Troponin T standard curve was prepared as described in Section 10. Raw data values are shown in the table. Background-subtracted data values (mean +/- SD) are graphed.

주요 정보

검출 방식

Colorimetric

샘플 타입

Cell culture media, Serum

Reacts with

Human

분석 유형

Sandwich

결과 유형

Quantitative

Sensitivity

= 68.2 pg/mL

검출 범위

195.3 - 12500 pg/mL

분석 소요 시간

1h 30m

분석 플랫폼

Pre-coated microplate (12 x 8 well strips)

Reactivity 정보

{ "title": "Reactivity Data", "filters": { "stats": ["", "Reactivity", "Dilution Info", "Notes"] }, "values": { "sELISA": { "reactivity":"TESTED_AND_REACTS", "dilution-info":"", "notes":"<p></p>" } } }

제품 세부 정보

Human Cardiac Troponin T ELISA Kit ab223860 is a rapid single-wash 90-min Sandwich ELISA to measure Human Cardiac Troponin T in cell culture media, serum. This SimpleStep sensitivity is 68.2 pg/mL.

How the assay works

Human Cardiac Troponin T SimpleStep ELISA®employs capture antibodies conjugated to an affinity tag that is recognized by the monoclonal antibody used to coat our SimpleStep ELISA® plates. This approach to sandwich ELISA allows the formation of the antibody-analyte sandwich complex in a single step, significantly reducing assay time. See the SimpleStep ELISA® protocol summary in the image section for further details.

Assay Specificity

Our SimpleStep ELISA® kits use recombinant monoclonal antibodies rigorously validated to ensure the highest level of consistency and reproducibility, improved sensitivity and specificity and ease of scalability and security of supply.
Please refer to our protocol booklet for more details.

Human Cardiac Troponin T ELISA Kit ab223860 protocol summary

1. Mix: add samples/standards to the wells together with the capture and detector antibody cocktail. Incubate 1 hr at room temperature
2. Wash
3. Add TMB development solution - incubate for 10 min
4. Add Stop solution
5. Read the results on a plate reader at 450 nm

Design your own immunoassay

We offer the antibody pair used in this kit in a BSA and Azide-free format, ready for conjugation:

- Anti-Cardiac Troponin T antibody [EPR20266-143] - BSA and Azide free (Capture) ab252655
- Anti-Cardiac Troponin T antibody [EPR20266-143-6] - BSA and Azide free (Detector) ab252655

Cardiac Troponin T is one of the three tropomyosin binding subunit that makes up the troponin complex with TnI and TnC being the other two subunits. The troponin complex is essential to the contraction of skeletal and heart muscles. Troponin T specifically binds to tropomyosin and helps position it on actin. Studies have shown elevated levels of troponin T after myocardial infarction and significantly associated with the incidence of heart failure and cardiovascular death.

REACH authorisation
Abcam has not and does not intend to apply for the REACH Authorisation of customers' uses of products that contain European Authorisation list (Annex XIV) substances.
It is the responsibility of our customers to check the necessity of application of REACH Authorisation, and any other relevant authorisations, for their intended uses.

Precision

[ { "reproducibilityType": "Inter", "sample": "Serum", "replicates": 3, "mean": null, "standardDeviation": null, "coefficientOfVariability": "1.96" }, { "reproducibilityType": "Intra", "sample": "Serum", "replicates": 8, "mean": null, "standardDeviation": null, "coefficientOfVariability": "4.82" } ]

Recovery

[ { "sample": "Serum", "range": "83 - 93 %", "average": "= 88" }, { "sample": "Cell culture media", "range": "94 - 98 %", "average": "= 96" }, { "sample": "Citrate plasma", "range": "80 - 89 %", "average": "= 84" }, { "sample": "Tissue Extracts", "range": "76 - 88 %", "average": "= 83" } ]

제품 구성

{ "values": { "1x96Tests": { "sellingSize": "1 x 96 Tests", "publicAssetCode":"ab223860-1x96Tests", "assetComponentDetails": [ { "size":"1 x 1 Unit", "name":"Plate Seal", "number":"AB223860-CMP09", "productcode":"" }, { "size":"1 x 6 mL", "name":"Antibody Diluent 4BI", "number":"AB223860-CMP08", "productcode":"" }, { "size":"1 x 1 Unit", "name":"SimpleStep Pre-Coated 96-Well Microplate (ab206978)", "number":"AB223860-CMP04", "productcode":"" }, { "size":"1 x 10 mL", "name":"5X Cell Extraction Buffer PTR (ab193970)", "number":"AB223860-CMP07", "productcode":"" }, { "size":"1 x 12 mL", "name":"Stop Solution", "number":"AB223860-CMP05", "productcode":"" }, { "size":"1 x 20 mL", "name":"10X Wash Buffer PT (ab206977)", "number":"AB223860-CMP11", "productcode":"" }, { "size":"1 x 50 mL", "name":"Sample Diluent NS (ab193972)", "number":"AB223860-CMP03", "productcode":"" }, { "size":"1 x 12 mL", "name":"TMB Development Solution", "number":"AB223860-CMP06", "productcode":"" }, { "size":"1 x 600 µL", "name":"10X Human Cardiac Troponin T Capture Antibody", "number":"AB223860-CMP01", "productcode":"" }, { "size":"2 x 1 Vial", "name":"Human Cardiac Troponin T Lyophilized Recombinant Protein", "number":"AB223860-CMP10", "productcode":"" }, { "size":"1 x 600 µL", "name":"10X Human Cardiac Troponin T Detector Antibody", "number":"AB223860-CMP02", "productcode":"" } ] } } }

특성 및 보관 정보

배송 시 보관 조건
Blue Ice
적절한 단기 보관 조건
+4°C
적절한 장기 보관 조건
+4°C
보관 정보
+4°C

제품 프로토콜

타겟 정보

Troponin T is the tropomyosin-binding subunit of troponin, the thin filament regulatory complex which confers calcium-sensitivity to striated muscle actomyosin ATPase activity.
See full target information TNNT2

대체 명칭 보기

TnTc, Cardiac muscle troponin T, cTnT, TNNT2

제품이 사용된 논문 (4)

Recent publications for all applications. Explore the 전체 목록 and refine your search

Advanced science (Weinheim, Baden-Wurttemberg, Germany) 11:e2401415 PubMed38965824

2024

Modeling the Effects of Protracted Cosmic Radiation in a Human Organ-on-Chip Platform.

Applications

Unspecified application

Species

Unspecified reactive species

Daniel Naveed Tavakol,Trevor R Nash,Youngbin Kim,Pamela L Graney,Martin Liberman,Sharon Fleischer,Roberta I Lock,Aaron O'Donnell,Leah Andrews,Derek Ning,Keith Yeager,Andrew Harken,Naresh Deoli,Sally A Amundson,Guy Garty,Kam W Leong,David J Brenner,Gordana Vunjak-Novakovic

Disease models & mechanisms 17: PubMed38050701

2024

Physiological stress improves stem cell modeling of dystrophic cardiomyopathy.

Applications

Unspecified application

Species

Unspecified reactive species

Dominic E Fullenkamp,Alexander B Willis,Jodi L Curtin,Ansel P Amaral,Kyle T Dittloff,Sloane I Harris,Ivana A Chychula,Cory W Holgren,Paul W Burridge,Brenda Russell,Alexis R Demonbreun,Elizabeth M McNally

Molecules (Basel, Switzerland) 28: PubMed37513325

2023

Phenolics Extracted from Mitigates Diabetic Cardiomyopathy by Modulating Oxidative Stress, Apoptotic Mediators and the Nfr-2/HO-1 Pathway in Alloxan-Induced Diabetic Rats.

Applications

Unspecified application

Species

Unspecified reactive species

Urooj Umar,Sibtain Ahmed,Asra Iftikhar,Maryam Iftikhar,Wafa Majeed,Atika Liaqat,Sana Shahzad,Mateen Abbas,Tahir Mehmood,Farooq Anwar

PLoS neglected tropical diseases 16:e0010864 PubMed36228040

2022

Clinical and prognostic relevance of sST2 in adults with dengue-associated cardiac impairment and severe dengue.

Applications

Unspecified application

Species

Unspecified reactive species

Andrew Teo,Po Ying Chia,Gaurav Kumar Ramireddi,Sebastian Kah Ming Khoo,Tsin Wen Yeo
제품이 사용된 논문 모두 보기
websiteProtocolBooklet
en

Product promise

저희는 고품질 시약으로 고객님의 연구를 서포트하고자 최선을 다하고 있으며, 모든 단계에서 함께 하겠습니다. 만약 제품이 기대한 성능을 보이지 않을 경우에도 Product Promise로 보호받으실 수 있습니다.
자세한 내용은 이용 약관을 확인해 주세요.

Please note: All products are 'FOR RESEARCH USE ONLY. NOT FOR USE IN DIAGNOSTIC OR THERAPEUTIC PROCEDURES'.

For licensing inquiries, please contact partnerships@abcam.com